Description
CD800a | Measuring range | Best accuracy | Resolution | Input impedance |
---|---|---|---|---|
DCV | 400m/4/40/400/600V | ±(0.7%+3) | 0.1mV | 10M~100MΩ |
ACV | 4/40/400/600V | ±(1.6%+5) | 0.001V | 10M~11MΩ |
DCA | 40m/400mA | ±(2.2%+5) | 0.01mA | |
ACA | 40m/400mA | ±(2.8%+5) | 0.01mA | |
Resistance | 400/4k/40k/400k/4M/40MΩ | ±(1.2%+5) | 0.1Ω | |
Capacitance | 50n/500n/5μ/50μ/100μF | ±(5%+10) | 0.01nF | |
Frequency | 5Hz~100kHz | ±(0.5%+3) | ||
Duty cycle | 20%~80% | ±(0.5%+5) | ||
Continuity | Buzzer sounds at between 10Ω and 120Ω Open voltage:Approx.0.4V | |||
Diode test | Open voltage:approx.1.5V |
Display | numeral display 4000 |
Sampling rate | 3 times/sec. |
Bandwidth | 40Hz~400Hz |
Fuse | 0.5A/250V 1.5kA Φ5×20 Ceramic |
Battery | R6P(1.5V)x2 |
Battery life | Approx.500h (manganese battery) at DCV range |
Size/Mass | H176×W104×D46mm/Approx.340g |
Test lead length/Pin tip | Length: approx. 0.7m/19mm |